Specifications
BASIC MEASUREMENTS
result taking into account the measurement error for the indication of alternating current intensity measurement: 8A = 8 ± 0.34A*
The table with the resolution and accuracy for a given measurement can be found in the instruction manual.
*The YT-73086 meter does not have the true rms measurement function, therefore the above result is the real effective value of the AC voltage only for the correct sine wave.
BUILT-IN FUNCTIONS
ADDITIONAL INFORMATION
- Weight (kg) 0.8130
- Master Carton MC 24
- Pal 240
- Auto range YES
- Frequency measurement 0-100kHz
- AC voltage measurement 0-600V
- DC voltage measurement 0-600V
- Capacitance measurement 0-100µF
- Resistance measurement 0-40MΩ
- Temperature measurement -20 to 1000°C
- Duty cycle measurement 5-95%
- AC 0-10A
- DC 0-10A
- Continuity test YES
- YES diode test
BASIC MEASUREMENTS
- Alternating voltage measurement: A.C. 0-600V
- DC voltage measurement: D.C. 0-600V
- Alternating current measurement: A.C. 0-10A
- DC current measurement: D.C. 0-10A
- Resistance measurement: 0-40MΩ
- Diode test: IF 1mA, UR 1.5V
- Continuity test: 0-50Ω
- Capacitor capacitance measurement: 0-100μF
- frequency measurement: 0-100khz
- temperature measurement: -20 to 1000°C
- duty cycle measurement: 5-95%
- Example of Accuracy Determination for AC Voltage Measurement. Calculations for the measurement result: 235V
- The measurement accuracy of the YT-73086 meter for this range is: ±(1.0% + 5), resolution 0.1V
- Calculation of measurement error formula: (measurement result x 1%) + 5 x resolution
- Calculation: (235 x 1%) + 5 x 1 = 2.35 + 0.5 = 2.85V
- Result taking into account the measurement error for the indication 235V = 235 ± 2.85V*
result taking into account the measurement error for the indication of alternating current intensity measurement: 8A = 8 ± 0.34A*
The table with the resolution and accuracy for a given measurement can be found in the instruction manual.
*The YT-73086 meter does not have the true rms measurement function, therefore the above result is the real effective value of the AC voltage only for the correct sine wave.
BUILT-IN FUNCTIONS
- AUTO - automatic adjustment of the measurement range. When measuring unknown values with a multimeter that is manually ranged using the dial, make measurements sequentially from the largest range to the range that gives the most accurate result - otherwise you risk damaging the multimeter. However, in the case of a multimeter with the AUTO function, it is enough to make one measurement and the meter will automatically find the optimal range.
- RANGE - manual change of the measurement range. A useful function when checking many of the same components with a known value, e.g. a series of resistors. Manual range setting will shorten the time of such measurements significantly.
- FUNC - changing the measurement mode, e.g. from AC to DC voltage measurement, switching the diode test to the circuit continuity test
- REL - using the function during the measurement will reset the display and adopt the value visible before the display as the reference level. The next measurement will show the difference between the measured value and the stored reference value.
- HZ% - the button in the position of the knob for frequency measurement will change the settings to the duty cycle measurement
- DATA HOLD - keeping the currently measured value on the display
ADDITIONAL INFORMATION
- LCD display, maximum displayed score: 3999
- screen backlight
- sound signal in the circuit continuity test for resistance below 50Ω
- Diode and continuity tests can be enabled separately
- the set includes measuring probes, a temperature measuring probe and a cover
- design with overload protection
- housing made of ABS plastic + so-called holster, i.e. a rubber casing protecting the meter against mechanical damage
- meter turns off automatically
- low battery indicator
- power supply: 9V(6F22) battery
Inquiry - YATO Digital Multimeter / Model: YT-73086